Reverse Osmosis Water Purification Systems

HDRO SR

Central reverse osmosis system for haemodialysis

The HDRO SR (touch screen control panel) supplies reverse osmosis water to haemodialysis machines and is used to prepare dialysate concentrate to AAMI water quality standards. It is designed to meet all current standards for dialysis quality water.

Features & benefits

Consistent & safe

  • Complies with AAMI and ISO 13485:2016
  • Registered with the Medical Device Authority (MDA)
  • Continuous alarm monitoring of raw and dialysis water conductivity
  • Continuous monitoring of all flow rates and pressures
  • Low inlet pressure alarm with pump shutdown
  • UV sterilisation (optional)

Innovative design

  • Space-saving design
  • LCD touch screen control panel
  • System alarm messages
  • Stainless steel cabinet
  • Water-saving system (optional)

User friendly

  • Easy-to-operate selector switch
  • Automatic backwash cycle
  • Sampling valves for soft and dialysis water

Technical specifications

ModelHDRO SR 10HDRO SR 15HDRO SR 20HDRO SR 25HDRO SR 30HDRO SR 35HDRO SR 40
Permeate out10 L/min15 L/min20 L/min25 L/min30 L/min35 L/min40 L/min
Reject out10 L/min15 L/min20 L/min25 L/min30 L/min35 L/min40 L/min
Rated capacity4,400 GPD6,600 GPD8,800 GPD11,000 GPD13,200 GPD15,400 GPD17,600 GPD
Membranes (units)2345678
Water pressure50–100 PSI
Power supply240 VAC / 50 Hz / 30 A or 415 VAC / 50 Hz / 20 A
Feed waterCity water, TDS < 10
System recovery50–75%
RO membranePolyamide thin-film composite
Salt rejection99.5% (ionic)
Water qualityDepends on inlet water quality. Designed to meet MOH guidelines and AAMI standards for haemodialysis; RO water complies with AAMI RD62.

Planning a new dialysis centre or upgrading your RO system?

Talk to our team for a recommendation and quotation tailored to your centre.

WhatsApp us